Key Benefits
✔ Effective Astigmatism Correction
Available in cylinder powers from T1 to T6 (1.00–4.50 D at the IOL plane), allowing for highly customized astigmatism correction tailored to each patient's specific needs.
✔ Superior Rotational Stability
-
Median rotation of only 2°..
-
100% of the IOLs exhibit a rotation of ≤ 4° from the end of surgery through 1-year post-op.
This ensures unwavering astigmatism correction and highly predictable refractive outcomes.
✔ Premium Glistening-Free Hydrophobic Material
Crafted from ultra-pure hydrophobic acrylic, delivering:
-
Superior optical transparency.
-
Proven absence of glistenings.
-
Zero adhesion between the optic and the haptics.
-
Extended durability and optimal optical quality.
✔ High-Order Aspheric Optics
Features an advanced toric aspheric design that minimizes optical aberrations and optimizes visual quality across varying lighting conditions.
✔ Sistema Prosert® 1-Step Preloaded System
-
Simple, highly controlled implantation.
-
Injector tip specifically engineered for micro-incisions.
-
Fully factory preloaded → minimizes handling and significantly reduces the risk of intraoperative complications.
✔ Engineered Capsular Stability
-
Features Modified L haptics equipped with "Tension Gaps."
-
360° square-edge design.
-
Precise alignment marks on the anterior surface to ensure accurate axis positioning.
TECHNICAL SPECIFICATIONS — For Ophthalmologists
Model:
Primus-HD® Toric / Primus-HD® Toric Yellow
AT1 – AT6
Material:
Hydrophobic acrylic, glistening-free.
-
Clear: Integrated UV filter.
-
Yellow: UV filter + blue light filter.
Optical Design:
-
Toric high-order aspheric
-
Convex Posterior Design
-
Axis alignment marks located on the anterior surface.
-
Optic Diameter: 6.00 mm
-
Overall Diameter: 13.00 mm
Rangos dióptricos:
-
Spherical: +5.00 D a +36.00 D (in 0.50 D increments).
-
Cylindrical: 1.00 D a 4.50 D (T1–T6)
-
T1: 1.00 D (0.72 D at corneal plane)
-
T2: 1.50 D (1.07 D at corneal plane)
-
T3: 2.25 D (1.59 D at corneal plane)
-
T4: 3.00 D (2.11 D at corneal plane)
-
T5: 4.00 D (2.80 D at corneal plane)
-
T6: 4.50 D (3.15 D at corneal plane)
-
Refractive Index:
1.48 (a 35°)
Haptics
Modified L-Loop
Angulation: 1.5°
Tension Gaps → enhanced stability and centration.
Edge Design:
360° continuous sharp edge (square-edge).
- Reduces Posterior Capsular Opacification (PCO).
- Improves the mechanical barrier against lens epithelial cells (LECs).
- Reduces edge glare via peripheral texturing.
Injector System:
Prosert® 1-Step Preloaded Injector.
-
Preloaded
-
Recommended for "in-the-bag" implantation.
Calculated Constants (Subject to surgeon optimization):
-
A Nominal: 118.9
-
Haigis: a0=1.499, a1=0.40, a2=0.10
-
Hoffer Q: pACD 5.688
-
Holladay: sf 1.949
-
SRK/T: 119.188
-
SRK II: 119.588






Reviews