Primus-HD® / Primus-HD® Yellow – Premium Hydrophobic Intraocular Lens

Serie:

Primus-HD® / Primus-HD® Yellow – Premium Hydrophobic Intraocular Lens

The Primus-HD® family of intraocular lenses is designed to deliver a clear, stable, and safe visual experience following cataract surgery or refractive lens exchange. Its ultra-pure hydrophobic acrylic material, advanced aspheric design, and fully preloaded injection platform ensure a predictable, rapid, and smooth implantation.

Primus-HD® está disponible en dos versiones:

  • Primus-HD® (Clear): A transparent, aspheric lens.

  • Primus-HD® Yellow: Incorporates a blue light filter (390–470 nm) for enhanced retinal protection without compromising visual transmission.

These lenses represent a premium solution for patients seeking outstanding visual clarity, surgical safety, and long-term capsular stability..

Benefits of the Primus-HD® IOL

✔ Premium Glistening-Free Hydrophobic Material

Its high-purity hydrophobic acrylate (Grade "0" on the Miyata scale) eliminates the occurrence of glistenings and delivers superior optical transmission.

✔ Next-Generation Aspheric Optics

Featuring a negative aberration correction of –0.20 μ, it significantly enhances image quality and contrast sensitivity, particularly in low-light conditions.

✔ Convex Posterior Design

Promotes excellent optical quality and ensures a highly stable relationship with the posterior capsule.

✔ 360° Square-Edge Design

Increases the mechanical barrier effect against lens epithelial cells (LECs), significantly reducing the risk of posterior capsular opacification (PCO).

✔ Prosert® 1-Step Preloaded Injector

  • 100% factory preloaded for a zero-touch technique.

  • Dynamic 1.78 mm tips, suitable for micro-incisions of 2.0–2.2 mm.

  • Rapid and highly controlled deployment.

  • Zero adhesion between the optic and the haptics.

✔ Exceptional Stability via Modified C-Loop

The 1.5° angled haptic design keeps the lens perfectly centered and stable, even in large capsular bags or with anatomical variations.


Primus-HD® Yellow – Added Protection

The Yellow version features an integrated blue light filter). (390–470 nm).
This helps reduce retinal exposure to high-energy light without altering natural chromatic perception or visual quality.


Optical Performance

  • High light transmission.

  • Textured edges to minimize unwanted reflections ("edge glare").

  • Excellent capsular adaptation with full 360° contact.


TECHNICAL SPECIFICATIONS — For Ophthalmologists

IOL Type:

Single-piece (monoblock), hydrophobic acrylic, glistening-free.

Optic:

  • Aspheric, aberration-correcting (–0.20 µm).

  • Convex posterior design.

  • Optic Diameter: 6.0 mm.

  • Overall Diameter: 13.0 mm.

Material:

Ultra-pure hydrophobic acrylate (Miyata "0").
Integrated UV filter.
(The Yellow version includes an additional blue light filter)..

Haptics

Modified C-Loop
Angulation: 1.5°

Edge Design:

360° square-edge with anti-glare peripheral texturing.

Dioptric Range:

–10.0 D a +36.0 D, in 0.5 D increments.
All powers are fully preloaded.

Refractive Index:

1.48 @ 35°.

Calculated Constants (Subject to surgeon optimization):

  • A-constant nominal: 118.9

  • Haigis: a0=1.46 , a1=0.40 , a2=0.10

  • Hoffer Q: pACD 5.68

  • Holladay: sf=1.91

  • SRK/T: 119.2

  • SRK II: 119.6

Injector / Implantation System:

System Prosert® Preload Injector

  • Dynamic design

  • 1.78 mm Tip

  • Recommended implantation: "In-the-bag"

  • Compatible incision: 2.0–2.2 mm

  • 3-step procedure: Wetting → Preparation → Implantation

Reviews

There are no reviews yet.

Be the first to review “Primus-HD® / Primus-HD® Yellow – Lente Intraocular Premium Hidrofóbica”

Your email address will not be published. Required fields are marked *

Frequently Asked Questions

Which patients are ideal candidates for the Primus-HD®?
It is indicated for patients with cataracts or those seeking refractive lens exchange with a premium monofocal lens that prioritizes visual clarity, long-term stability, and a glistening-free optic.
No. The blue light filter is carefully optimized to protect the retina from high-energy light without altering natural color perception or visual quality.
  • Enhanced contrast and sharpness due to aspheric aberration correction.
  • Premium glistening-free hydrophobic material.
  • Advanced edge architecture specifically designed to minimize posterior capsular opacification (PCO).
  • Safer, zero-touch implantation as the system comes 100% preloaded.
The Modified C-Loop design and 1.5° haptic angulation provide excellent centration and exceptional long-term capsular stability, significantly reducing the risk of postoperative displacement.
It follows the standard postoperative care protocols for cataract surgery, including the use of prescribed anti-inflammatory eye drops, maintaining proper ocular hygiene, and avoiding strenuous activities during the initial recovery period.

How can we help you?

Complete the form below and an EverMED specialist will get in touch with you shortly.

Location

Av. Máximo Gómez No. 60, Suite 220, La Esperilla, Santo Domingo, D.N.

Phone

+1 809-620-5415

Email

info@evermedrd.com

WhatsApp / Chat

Available during business hours

Request Info or Quote

Ready to Elevate Your Standard of Care?

Contact us today, and an EverMED specialist will guide you toward the best solutions for your practice.
Scroll to Top
Complete the form below and an EverMED specialist will get in touch with you shortly.